Clinical Study of IOL Master for Measurements of Axial Length and Corneal Curvature before and after Removal of Silicone Oil
Author: XiaoJing
Tutor: MaXiang
School: Dalian Medical University
Course: Ophthalmology
Keywords: IOL Master Axial Curvature of the cornea Silicone oil
CLC: R779.6
Type: Master's thesis
Year: 2010
Downloads: 76
Quote: 0
Read: Download Dissertation
Abstract
|
Objective: To compare silicone oil removal before and after surgery IOL Master and other commonly used clinical instrument accuracy of axial length and corneal curvature measurement , to investigate its clinical characteristics . Methods : analysis of 34 cases cataract extraction plus vitrectomy combined with silicone oil tamponade in patients with axial length measurements using the IOL Master and A , respectively, before and after silicone oil removal , the IOL Master corneal curvature measuring the curvature of the cornea . Results : 1.IOL Master measurement silicone oil removal preoperative axial length was 24.18 ± 2.11 mm and postoperative axial length was 24.12 ± 1.98mm , A ultra measurement silicone oil remove preoperative axial length was 32.77 ± 3.41 mm axial growth after 24.33 ± 2.03mm. The measuring results of the IOL Master axial length before and after surgery the difference was not statistically significant ( P = 0.369 ) A super- axial length measurements before and after surgery , a significant difference ( P lt; 0.01 ) ; preoperative two instruments the measurement results are significant differences (P lt; 0.01), and the measurement results of the two instruments after the difference was not statistically significant ( P = 0.062) . 2.IOLMaster measurement silicone oil removal preoperative corneal curvature of 42.81 ± 2.41 D, corneal curvature to 42.89 ± 2.32D ; keratometer measured pre-operative corneal curvature was 42.46 ± 2.13D , corneal curvature was 42.55 ± 2.12D . The IOL Master keratometer There was no statistically significant difference in corneal curvature measurements before and after silicone oil removal . Silicone oil removal preoperative the IOL Master keratometer measurements mean difference -0.35D ( keratometer -IOL Mastr ) , the 95 % limits of agreement for -2.49 to 1.79D ; poor results after silicone oil removal both measurement the value of the mean -0.34D ( keratometer -IOL Mastr ) , the 95 % limits of agreement for -2.37 to 1.69D . Conclusion : For silicone oil removal before and after the axial length measurement the IOL Master 's accuracy is higher than supine contact A ??super : For the measurement of the curvature of the cornea , IOL Master more accurate than keratometer .
|
Related Dissertations
- Development of the Platform for Compressor Optimization Design and Aerodynamic Optimization Design in the Transonic Compressor,TH45
- Development of the On-Line Electronic Gauge for Crackshaft’s Axial Length Parameters,TH822
- Risk Factors for the Development of Transient Hypotony after Sinlicone Oil Removal in Aphakia Eyes,R779.6
- Correlation of Branch Retinal Vein Occlusion and Ocular Axial Length、 Anterior Chamber Depth,R774.1
- The Analysis of the Growth Factor Related with the Secondary Glaucoma in the Anterior Champer of Silicone Oil-filled Eyes,R779.6
- The Clinical Evaluation of Measuring the Horizontal Sulcus to Sulcus Distance by Ultrasound Biomicroscopy,R770.4
- Mechanical Property Analysis on Concrete Short Columns with Strengthened Style Hoop,TU375.1
- The Preparation of Modified Siloxane Surfactant and Its Application in Emulsifying Silicone Oil,TQ423
- Study on Yield Mechanism of Reinforced Concrete Frame Structure’s Strong Column-Weak Beam under Rare Earthquake Action,TU275
- No leakage electric pump Loss Analysis and axial force balance new method for research,TH38
- Field Test Research on the Vertical Load-bearing Characteristics of Bridge Pile Foundation in Steep-slope Area of Qinba Mountain Region,U443.15
- Clinical Application of IOL Master in Patients before and after Laser-assisted in Situ Keratomileusis,R779.63
- Design and Synthesis Axially Chiral Phosphiamines Ligand by Using a Stereocontrol System,TQ203.2
- The Experimental Investigation and Nonlinear Finite Element Analysis of Z-Shaped Column’s Shearing Performance Subjected to Low-Cyclic Horizongtal Load,TU375.3
- Influence of Column Axial Force to Weld Internal Force at Beam-to-column Joint in All-welded Steel Frame,TU391
- In Vitro Study on the Effects of Different Medicine to Silicone Oil Emulsification,R779.6
- Study on Dynamic Modeling and Boundary Identification of Axial Compression Column Under Different Pressures,TU311.3
- IOL Master biometry and IOL power calculation clinical observation,R318.1
- Numerical Simulations of Solute Taylor Dispersion in Capillary,TQ021.1
- CFST stub columns of Carrying Capacity,TU398.9
- GFRP reinforcement poles Experimental study of axial compression,TU366.2
CLC: > Medicine, health > Ophthalmology > Eye surgery and surgery
© 2012 www.DissertationTopic.Net Mobile
|