Application of ORA in the study of biomechanical properties of the cornea.

Author: QiuJiaJi
Tutor: LiWeiYe
School: Beijing Union Medical College
Course: Clinical
Keywords: ocular response analyzer introcular pressure cornealhysteresis corneal resistance factor glaucoma kerato-refractive surgery
CLC: R770.4
Type: PhD thesis
Year: 2010
Downloads: 22
Quote: 0
Read: Download Dissertation

Abstract


Objectives:To approach the accuracy and reliability of introcular pressure (IOP) measured by Reichert Ocular Response Analyzer (ORA),to compare and analyze the relationship between central corneal thickness (CCT) and IOP value measured by ORA,Goldmann applanation tonometry (GAT) and Non Contact Tonometer (NCT).Then to study the clinical utility of the corneal hysteresis (CH) and corneal resistane factor (CRF) measured by ORA in normal, glaucoma patient and suspect, undergone kerato-refractive surgery subjects.Method:Chose84healthy individuals (168eyes) for IOP measuring from November2009to March2010randomly. IOP was measured by ORA and NCT devices,then CCT was measured by A-ultrasond (A-US). Then compared the measuring values of these tonometries with each other and correlated them with CCT seperately. Also, Chose31glaucoma and suspect (62eyes) for IOP measuring. IOP was measured by ORA and GAT devices, then CCT was measured, by A-ultrasond (A-US). In addition, chose other51individuals (102eyes) undergone kerato-refractive surgery to this study. IOP was measured by ORA and NCT devices, then CCT was measured by A-ultrasond (A-US). CH, CRF, IOPg and IOP corneal-compensated (IOPcc) were provided by the ORA device. Then use SPSS to analyze the data.Results:The normal intraocular pressure values, ORA Goldmann IOP value measured analog (Goldmann-correlated IOP value, IOPg) was17.90±4.17mmHg, corneal compensated IOP (Corneal-Compensated Intraocular Pressure, IOPcc) was18.34±4.00mmHg; NCT IOP value was17.44±3.54mmHg. IOPcc, IOPg, NCT was good correlation between intraocular pressure, the mean value between each were significantly different. CCT was536.20±29.17μm. NCT IOP measurements and CCT medium level of the correlation coefficient was0.429; and IOPcc shows only low levels of weak correlation between CCT, the correlation coefficient was0.258. Significantly higher than normal male IOPg to female, IOPcc no significant difference, IOPcc, IOPg no correlation with age. Normal CH and CRF were 10.13±2.11mmHg and10.90±2.31mmHg, with age, gender was no significant correlation. CH and CCT has low correlation coefficient of0.277; CRF and CCT medium level of the correlation coefficient of0.466.Glaucoma and suspected glaucoma IOPg was19.16±4.10mmHg, IOPcc was19.91±4.44mmHg; GAT IOP value of18.91±3.75mmHg. Glaucoma and suspected glaucoma IOPg, IOPcc, GAT correlation among good, average significantly higher than among IOPcc and IOPg, GAT IOP. There is no significant difference between GAT IOP values and IOPg. GAT IOP and CCT has medium level correlation and the correlation coefficient of0.344. Patients with glaucoma and suspected glaucoma CH and CRF were9.83±2.13mmHg and11.03±2.11mmHg. Compared with the normal group, glaucoma and suspected glaucoma IOPg, IOPcc significantly higher than the normal group, and CH, CRF, CCT was no significant difference.After corneal refractive surgery, LASEK patients were followed up by IOPg was13.26±3.52mmHg, IOPcc was16.64±3.29mmHg, NCT IOP value was14.13±3.61mmHg, CH was8.01±1.60mmHg, CRF was7.70±1.82mmHg, CCT was447.59±43.75μm; LASIK patients were followed up by IOPg was13.10±5.29mmHg, IOPcc was16.70±5.25mmHg, NCT IOP value was13.91±4.99mmHg, CH was7.83±1.52mmHg, CRF was7.50±1.75mmHg, CCT was447.96±46.22μm. Compared with normal subjects, patients after corneal refractive IOPg, IOPcc, NCT IOP, CCT, CH, CRF were lower, but IOPcc decreased no significantly than NCT IOP. Between LASEK surgery group and LASIK surgery group,IOPg, IOPcc, NCT IOP, CCT, CH, CRF were no significant difference.Conclusion:ORA IOP measurement with high repeatability, It has good consistency with GAT, NCT and less impact with the measurement of central corneal thickness. ORA CH and CRF measurement parameters can reflect the biomechanical properties of the cornea. Normal CH, CRF value has no significant difference between men and women, and age but low levels of correlation with CCT. In glaucoma and suspected glaucoma patients, CH, CRF value has no significant difference compared to normal controls. The CH value in patients afler corneal refractive surgery was significantly lower than the normal value and also the CRF. But there was no significant difference between LASEK surgery group and LASIK surgery group. ORA IOP measurements may be more closer to the real value than GAT, NCT value, especially after corneal refractive surgery.

Related Dissertations

  1. The Correlation Study between Entothelin-1 Level in Blood Plasma and Hemorrheology Changes in Primary Glaucoma and Visual Function Damage Degree.,R775
  2. A Clinical Study of Assessing the Optic Pathway Changes of Glaucoma by Magnetic Resonance Imaging,R775
  3. Study on the Correlation btween Serum, Aqueous Humor VEGF, IL-6 Levels and the Neovascular Glaucoma,R775
  4. The Analysis of the Growth Factor Related with the Secondary Glaucoma in the Anterior Champer of Silicone Oil-filled Eyes,R779.6
  5. OCT in early diagnosis of primary glaucoma clinical application of,R775
  6. Study of the Visual Pathway with MRI on Patients with Primary Glaucoma,R775
  7. Glaucoma Patient’s Psychological Feature and Psychological Intervention,B849
  8. The Effects and Underlying Mechanisms of S-allyl L-cysteine on the Retina Injury Subjected to Ischemia/reperfusion,R962
  9. Correlation Analysis of the structure and function of glaucoma,R775
  10. The Expression of TNF-α on Retinal Excitotoxicity Injury Induced by NMDA,R774.1
  11. Acute angle-closure glaucoma after phacoemulsification corneal endothelial cell injury and related influencing factors,R779.66
  12. The Study of Artificial Meshworks Made from Three-dimensional Porous Ceramics in Rabbits,R779.6
  13. Expression of Thymosin Beta4 in Human Tenon’s Capsule Fibroblasts and the Effect of CTGF on Its Expression,R779.6
  14. Association between Genetic Polymorphisms of the FP Receptor (Prostaglandin F2α Receptor) Gene and Response to Latanoprost,R775
  15. Analysis of Surgical Effect of Two Different Operations for Neovascular Glaucoma,R779.6
  16. Risk Factors of Shallow Anterior Chamber after Trabeculectomy in Primary Angle Closure Glaucoma,R779.6
  17. The Utilization of the Realtime Fourier-domain Anterior Segment OCT (FD-OCT) in the Image and meVinsanteurement of Schlemm’s Canal and the Diagnosis of Primery Open Angle Glaucoma,R775
  18. Experimental Study of CO2 Laser Assisted Filtering Surgery on Rabbits Intraocular Hypertension Model,R779.6
  19. Study on Self-management and Visual-related Quality of Life among Patients with Glaucoma,R473.77
  20. The Relationship between FD-OCT Structure and Visual Field Function in Glaucoma,R775
  21. The Study on Corneal Biomechanical Property for Normals and Eyes with Corneal Refractive Surgery by Using Ocular Response Analyzer (ORA),R779.63

CLC: > Medicine, health > Ophthalmology > Eye diagnostics
© 2012 www.DissertationTopic.Net  Mobile